Applications
● Blood glucose meter
● Blood oxygen meter
● COVID-19 PCR test machine
● Household beauty device
● Physiotherapy equipment
Description
ADA540K180S001B is a desktop power adapter with a single output. With 100 -240 VAC universal input voltage, it can continuously output any DC voltage of 18VDC, 1M output cable length. Widely used in portable medical equipment .
Naming Convention
ADA540K180S001B: Output Voltage 18V, Output Power 54W, medical desktop power adapter AC / DC |
||
□□ |
Product Code |
AD=Isolated AC/DC; DD=Isolated DC/DC |
□ |
Packaging Form |
A=adapter; M=resin filled power module; P=PCB module |
XXX |
Power Coding |
First two digits multiplied by 10 to the power of the third digit |
□ |
Application Level |
C=commercial grade T=industrial grade K=medical grade S=harsh environment application |
XXX |
Output Voltage |
*10-1V |
□ |
Custom Tag |
S=standard product Z=customized product |
XXX |
Product ID |
001-999, used to identify different products of the same category |
□ |
Major Upgrade Information |
A-Z, fixed as A for the first mass production |
Input Parameter
Voltage Range |
100~240 VAC |
Frequency Range |
50~60Hz |
Input Current (Max.) |
1.0A @ 115VAC,0.45A @ 230VAC |
Efficiency (Typ.) |
89% |
Standby Consumption |
0.2W |
Impulse Current (Typ.) |
6.6A @ 115VAC,5A @ 230VAC |
Leakage Current (Typ.) |
100uA RMS @ 264VAC/60Hz |
Output Parameter
Output Voltage |
18VDC |
Voltage Tolerance |
±5% |
Output Current |
3000mA |
Rated Power (Max.) |
54W |
Line Regulation |
±1% at full load |
Load Regulation |
±3% |
Ripple & Noise |
120mV (Typ.), 150mV (Max.)@20MHz |
Frequency (Typ.) |
65kHz |
Hold Up Time (Typ.) |
0.65mS @ 115VAC 0.7mS @ 230VAC |
Environment
Operating Temperature |
-10 ~ +40℃ |
Storage Temperature |
-20 ~ +85℃ |
Storage Humidity |
10% ~ 95%RH |
Soldering Temperature |
260±5℃ @ Wave Soldering, 5~10s 360±10℃ @ Manual Soldering, 3-5s |
Protection Function
Short Circuit |
Long term short circuit, auto recovery. |
Over Current |
≥130%IO auto recovery |
Over Voltage |
≤18.9VDC @ 18V output |
IEC Safety Class |
CLASSII |
Electric Shock Protection |
2×MOPP @ primary to secondary |
Reliability
MTBF |
≥30,000H @ 25℃ |
Safety Standards / Directives
Medical Safety |
EN60601-1 |
|
CE |
Compliant |
|
Isolation Voltage (Tpy.) |
Input to Output |
4000VAC @ 1 minute test and the leakage current is smaller than 5mA. |
EMC
Parameter |
Standard |
Test Level / Note |
Conducted Emission |
CISPR 11:2015+A1:2016+A2:2019 |
CLASS B |
Radiated Emission |
CISPR 11:2015+A1:2016+A2:2019 |
CLASS B |
Electrostatic Discharge |
IEC 61000-4-2:2008 |
±8KV/Contact ±15KV/Air |
Radiate Susceptibility |
IEC 61000-4-3:2006+A1:2007+A2:2010 |
10V/m |
Electrical Fast Transient Burst |
IEC 61000-4-4:2012 |
±2 KV |
Surge |
IEC 61000-4-5:2014+A1:2017 |
±1KV (line to line) |
Conducted Susceptibility |
IEC 61000-4-6:2013 |
10Vr.m.s |
Voltage Dips and Interruption |
IEC 61000-4-11:2004+A1:2017 |
0%, 70% |
Mechanical Specification
Dimension |
126.00*50.00*30.30(H)mm |
Enclosure Material |
Black flame retardant and heat resistant plastics(UL94V-0) |
Cooling Mode |
Natural air cooling |
Power and Cable Specification |
ul2464×20Awg*1m,5.5*2.5*straight head+magnetic ring+SR |